Abstract
Introduction. Drug-coated balloon (DCB) angioplasty is a “leave-nothing-behind” strategy for selected coronary lesions, avoiding permanent metallic implantation. However, extensive lesion preparation and balloon oversizing may cause vessel-wall injury, including dissection and intramural hematoma. This report describes the successful management of acute coronary occlusion caused by an intramural hematoma during DCB angioplasty, highlighting cutting-balloon fenestration as a bailout strategy.
Methods. A significant right coronary artery lesion was prepared using a 3.0 × 12 mm non-compliant balloon inflated to 12 atm, followed by a 3.5 × 15 mm paclitaxel-coated DCB inflated to 8 atm. Optical coherence tomography (OCT) was performed to assess vessel-wall response.
Results. Pre-DCB OCT showed no vessel-wall injury. Following DCB inflation, OCT revealed an evolving intramural hematoma with associated dissection and progressive true-lumen compression, ultimately causing complete coronary occlusion. DCB inflation and possible oversizing were considered contributing factors. A 3.0 × 12 mm cutting balloon inflated to 12 atm created controlled fenestrations, decompressing the hematoma and immediately restoring coronary flow. Given the extent of vessel-wall injury, a 3.0 × 18 mm drug-eluting stent was subsequently implanted at 12 atm, achieving final TIMI 3 flow.
Discussion. Meticulous lesion preparation and appropriate balloon sizing are essential for safe DCB angioplasty. In cases of occlusive intramural hematoma, conventional balloon dilatation may be insufficient to achieve decompression. Cutting-balloon fenestration may restore coronary flow and facilitate definitive stent implantation.

This work is licensed under a Creative Commons Attribution 4.0 International License.
Copyright (c) 2026 Dona Zarevska, Biljana Zafirovska Taleska, Aleksandar Jovkovski, Ivan Vasilev, Hajber Taravari, Danica Petkoska Spirova, Bozin Shopov, Nikola Manev, Jorgo Kostov, Sasko Kedev